Gas Leak Detectors: Sensor Types and Buying Guide

Illustration of a four-gas detector showing LEL, O2, CO and H2S readings

Gas detection covers two different jobs. One is finding where a leak is. The other is warning people when the air around them becomes dangerous. The instruments overlap but are not interchangeable, and buying the wrong kind is a common mistake. This guide explains the sensor technologies, the instrument types and the specifications to check, for plant maintenance teams, HVAC and gas engineers, and safety managers.

Leak detector or gas monitor

  • A leak detector has a probe, often on a flexible neck, and a fast, sensitive sensor. You sweep it along joints, valves and fittings to pinpoint a leak. It is a diagnostic tool.
  • A personal or area gas monitor measures concentration continuously and alarms at set thresholds. It is worn or placed in a work area as safety equipment, and is typically certified for use in hazardous atmospheres.

If the purpose is protecting people in confined spaces or process areas, you need a certified gas monitor, not a leak sniffer.

Sensor technologies

Sensor Detects Notes
Catalytic bead (pellistor) Combustible gases, shown as percent of the lower explosive limit Needs oxygen to work. Can be poisoned by silicones and some other compounds
Electrochemical Oxygen and toxic gases such as carbon monoxide and hydrogen sulphide Gas-specific. Sensors have a limited life and are replaced periodically
Infrared (NDIR) Hydrocarbons and carbon dioxide Works without oxygen and resists poisoning. Does not detect hydrogen
Photoionisation (PID) Volatile organic compounds at low concentrations Very sensitive but not specific to one compound
Semiconductor Combustible gases and refrigerants, in handheld sniffers Sensitive and inexpensive. Best for locating leaks, not for measurement
Ultrasonic The sound of gas escaping under pressure Finds leaks of any gas, including compressed air. Does not measure concentration

Single-gas or multi-gas

A single-gas monitor covers one hazard, such as carbon monoxide. A multi-gas monitor carries several sensors. The common four-gas configuration covers combustible gas, oxygen, carbon monoxide and hydrogen sulphide, which matches the hazards assessed before confined space entry. Five and six-gas instruments add a PID or further toxic sensors.

Specifications to check

  • Target gases and ranges. Match the sensor set to the hazards in your risk assessment.
  • Hazardous area certification. For use where flammable gas may be present, the instrument must be certified intrinsically safe under the scheme your region uses, such as ATEX, IECEx or the North American equivalents.
  • Diffusion or pumped. Diffusion instruments sample the air around them. A pump lets you draw a sample from a space before entering it.
  • Response time. Quoted as the time to reach 90 percent of the final reading. Faster is better for leak tracing.
  • Alarms. Audible, visual and vibrating alarms, with low, high, short-term and time-weighted thresholds for toxic gases.
  • Battery life. It should cover a full shift with margin, including pump use.
  • Data logging. Useful for incident records and compliance.
  • Ruggedness. Ingress protection and drop resistance for site use.

Testing and calibration

Gas sensors drift and can fail without any visible sign. Two routines keep an instrument trustworthy:

  • Bump test: a brief exposure to test gas to confirm the sensors respond and the alarms work. Many safety programmes require it before each day's use.
  • Calibration: adjusting the readings against a certified gas concentration, at the interval the manufacturer specifies.

Factor the cost of calibration gas, a docking or test station and replacement sensors into the purchase.

Browse our inspection instruments for professional detection and measurement equipment.

Frequently asked questions

What does percent LEL mean?

It is the gas concentration as a percentage of the lowest concentration that can ignite in air. An alarm at 10 percent LEL warns you well before the atmosphere becomes flammable.

How long do the sensors last?

Electrochemical sensors commonly last two to three years. Catalytic and infrared sensors usually last longer. The instrument reports when a sensor is failing calibration.

Can one instrument detect every gas?

No. Each sensor responds to particular gases, so the sensor set has to match your hazards.

Do I need a pumped instrument?

You do if you need to test a space before entering it. For personal monitoring in open areas, diffusion is sufficient.